本平台为互联网非涉密平台,严禁处理、传输国家秘密或工作秘密

基于2,7-二(4-吡啶基)吖啶的荧光探针灵敏检测烟草农药残留抑芽丹

Sensitive detection of pesticide residue maleic hydrazide in tobacco using a fluorescent probe based on 2, 7-di(4-pyridyl)acridine

  • 摘要: 为实现烟草农药残留抑芽丹的快速定量检测,利用实验室自行设计合成的有机小分子荧光探针2, 7-二(4-吡啶基)吖啶与抑芽丹分子间存在荧光作用机制,建立了烟草中抑芽丹的快速检测方法。结果表明:①抑芽丹在0.5~100.0 μmol/L范围内线性关系良好,检出限为0.1 μmol/L(0.11 mg/kg)。②该方法荧光响应时间为1 min,测定快速。③10倍浓度的其他农药对待测液的荧光强度影响不大,方法选择性好;将其用于烟草中农药残留抑芽丹检测,加标回收率介于91.0%~106.7%之间,相对标准偏差为1.4%~4.8%。该方法适用于烟草中农药残留抑芽丹的快速定量检测。

     

    Abstract: In order to rapidly and quantitatively detect pesticide residue maleic hydrazide (MH) in tobacco, a rapid detection method was established by using the fluorescence between 2, 7-di(4-pyridyl)acridine, a small bespoke organic molecule fluorescent probe and MH molecule. The results showed that: 1) The linearity was satisfactory in the range of 0.5-100.0 μmol/L of MH, and the limit of detection was 0.1 μmol/L (0.11 mg/kg). 2) This method was rapid with the fluorescence response time of 1 min. 3) The fluorescence intensity of the tested solution had less interference by other pesticides at 10-fold concentrations, so the selectivity was high. The spiked recoveries ranged from 91.0% to 106.7% with relative standard deviations of 1.4%-4.8%. This method is thus suitable for the rapid and quantitative determination of pesticide residue MH in tobacco.

     

/

返回文章
返回