Abstract:
To reveal the influence patterns of motor frequency, internal temperature, equivalent contamination degree, and lamp voltage on the measurement error of an infrared moisture analyzer, response surface methodology was introduced to conduct a multi-factor coupling analysis of the operating parameters of the infrared moisture analyzer. Through a full-factorial experimental design, the main effects and interaction effects of each factor were quantitatively separated, thereby overcoming the limitation that traditional single-factor experiments cannot reveal coupling effects among factors. The results showed that: 1) Motor frequency has a relatively small effect on the measurement results, while internal temperature, equivalent contamination degree, and lamp voltage are the main operating parameters affecting the measurement results of the moisture analyzer; 2) In the multi-factor coupling experiments, lamp voltage and equivalent contamination degree are the main influencing factors, and internal temperature is a secondary influencing factor, with contribution rates to the measurement error of 44.94%, 42.70%, and 11.37%, respectively, while the contribution rates of all interaction terms do not exceed 0.18%; 3) Based on the response surface model and safety domain analysis, a quantitative compensation formula was inversely derived, enabling the transformation of the compensation strategy from post-event maintenance to pre-event warning. When the equivalent contamination degree reaches 36% and the internal temperature is below 47 ℃, the measurement error cannot be returned to the safe region by adjusting the operating parameters. This study quantified the effects of the instrument’s internal temperature, equivalent contamination degree, and lamp voltage on moisture measurement error, and provides technical support for the formulation of compensation strategies and the condition assessment of infrared moisture analyzers.