Abstract:
In order to improve the accuracy of test results, the uncertainty in the determination of benzoic acid in tobacco casing by HPLC was systematically calculated according to JJF 1059-1999
Evaluation and Expression of Uncertainty in Measurement. The results indicated that: 1) calibration curve and recovery of benzoic acid were the main factors influencing the uncertainty;2) the combined uncertainty was 0.23 mg/kg and the expanded uncertainty was 0.46 mg/kg (level of confidence 95%,