SHI JIE, LI LI, HU QING-YUAN, CHEN ZAI-GEN, WANG FANG. Simultaneous Determination of Cr, Ni, As, Se, Cd, Hg and Pb in Tobacco with ICP-MSJ. Tobacco Science & Technology, 2006, 39(12): 29-34,37. DOI: 10.3969/j.issn.1002-0861.2006.12.008
Citation: SHI JIE, LI LI, HU QING-YUAN, CHEN ZAI-GEN, WANG FANG. Simultaneous Determination of Cr, Ni, As, Se, Cd, Hg and Pb in Tobacco with ICP-MSJ. Tobacco Science & Technology, 2006, 39(12): 29-34,37. DOI: 10.3969/j.issn.1002-0861.2006.12.008

Simultaneous Determination of Cr, Ni, As, Se, Cd, Hg and Pb in Tobacco with ICP-MS

  • The Cr, Ni, As, Se, Cd, Hg and Pb in 85 tobacco samples from five major production areas of Chinawere simultaneously determined by inductively coupled plasma mass spectrometry (ICP-MS) with microwave digestion.The results showed that:1) the recoveries of these heavy elements were in the range of 96.34% to 106.7% with the RSD of 1.41%-9.78% and the limits of detection were from 7.974-147.8ng/L;2) the contents of these elements between different cultivars and production areas differed significantly, while those between different stalk positions did not for the same cultivar.
  • loading

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return