LEI MIN, LI SHAO-YE, YANG CHUN-PING, LIU LI-QING, XIE GENG-XIN. Selection of Microwave Digestion Conditions for Determination of Arsenic Content in Tipping PaperJ. Tobacco Science & Technology, 2010, 43(4): 44-47,64. DOI: 10.3969/j.issn.1002-0861.2010.04.010
Citation: LEI MIN, LI SHAO-YE, YANG CHUN-PING, LIU LI-QING, XIE GENG-XIN. Selection of Microwave Digestion Conditions for Determination of Arsenic Content in Tipping PaperJ. Tobacco Science & Technology, 2010, 43(4): 44-47,64. DOI: 10.3969/j.issn.1002-0861.2010.04.010

Selection of Microwave Digestion Conditions for Determination of Arsenic Content in Tipping Paper

  • In order to accurately determine As in tipping paper samples by HG-AFS with microwave digestion,the effects of digestion system,digestion procedure,acid removal and type of microwave instrument on As determination were investigated.The results showed that:1) tipping paper samples were completely digested in HNO3:HF:HCl:H2O2=6:0.2:1:1(v/v) system in a microwave field;2) the recoveries of As was from 98% to 110% with RSD of 6% to 10% and detection limit of 1.38 ng/L;3) the performance of microwave instrument Berghof was virtually the same as that of CEM.